MAINTENANCE
SECTION 3
17
Cleaning the X12+ and Accessories
1.
Remove cables and disconnect power source from device before cleaning.
2.
Wash the carry case by hand with fabric detergent and then air dry. Do not machine dry the case.
3.
For general cleaning, use a soft, lint-free cloth lightly moistened with a mild soap and water solution. Wipe and
air dry.
Do use clean, lint-free cloth
Don’t use alcohol or solvents
Don’t use abrasive cleaners or materials
4.
For disinfecting, wipe exterior with a soft, lint-free cloth using a solution of Sodium Hypochlorite (10%
household bleach and water solution) minimum 1:500 dilution (minimum 100 ppm free chlorine) and maximum
1:10 dilution as recommended by the APIC Guidelines for Selection and Use of Disinfectants.
5.
Use caution with excess liquid as contact with metal parts may cause corrosion.
6.
Do not immerse cable ends or lead wires; immersion can cause metal corrosion.
7.
Do not use excessive drying techniques such as forced heat.
WARNING: Prevent liquid from penetrating the device and do not attempt to clean/disinfect the device or
patient cables by submerging into a liquid, autoclaving, or steam cleaning. Never expose cables to strong
ultra-violet radiation. Do not sterilize the device or ECG cable with Ethylene Oxide (EtO) gas.
Periodic Maintenance
Check the X12+ and the ECG cable before each use to ensure they are not damaged or broken.
1.
Patient Cable Maintenance:
•
Check patient cables for cracks or breakage prior to use
•
Clean the cable with a germicidal solution that does not contain alcohol
•
Alcohol will cause hardening and can introduce cracks
•
Don’t use tape on the patient cable; tape residue will cause hardening and can introduce cracks
•
Patient cables should be stored by looping them loosely. Don’t pull or stretch the cables; don’t
wrap cables tightly
•
Replace patient cables periodically (depending on use and care)
2.
Exterior Visual Inspection:
•
Check connectors for loose, bent, or corroded contact points
•
Inspect covers for warping, surface damage, or missing hardware
•
Check for any other form of damage
Disposal of Waste Materials
The X12+ needs one alkaline battery and disposable monitoring electrodes. Disposal must be in accordance with
the following procedures:
Battery: applicable disposal or recycling standards
Electrodes: normal waste