SECTION 2
15
Setting the Number of Patient Cable Lead Wires
CABLE
is used to set the number of lead wires for the patient cable.
From the
CONFIGURE
menu, use
Down
or
Up/Right
to scroll to
CABLE
. Press
Enter
to select. The
CABLE
menu includes:
4 lead wire
5 lead wire
10 lead wire
Use
Down
or
Up/Right
to scroll to the desired option and press
Enter
. When finished, press
Enter
to exit the
CABLE
menu.
NOTE: When a 4-wire or 5-wire cable is selected, only the limb leads (and a V label with the 5-wire
cable) will appear in
LEAD CHECK
,
DISPLAY ECG,
and lead fail messages.
Setting Language
LANGUAGE
is used to select the language displayed in the main menu and all sub-menu options.
From the
CONFIGURE
menu, use
Down
or
Up/Right
to scroll to
LANGUAGE
. Press
Enter
to select. When
finished, press
Enter
to exit the
LANGUAGE
menu.
Viewing Software Version Number
VERSION
displays the current software version installed in the X12+.
From the
CONFIGURE
menu, use
Down
or
Up/Right
to scroll to
VERSION
. Press
Enter
to select and view the
current software. When finished, press
Enter
to exit the
VERSION
menu.
Viewing Battery Voltage
BATTERY
displays the voltage of the battery currently installed in the X12+.
From the
CONFIGURE
menu, use
Down
or
Up/Right
to scroll to
BATTERY
. Press
Enter
to select and view the
current battery voltage. When finished, press
Enter
to exit the
BATTERY
menu.