![background image](http://html1.mh-extra.com/html/mortara/quinton-cr60/quinton-cr60_user-manual_1836792021.webp)
13
8. ELECTROMAGNETIC COMPATABILITY (EMC)
Table X-1 Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment: Guidance
RF Emissions
CISPR 11
Group 1
The CR60 Rehabilitation Treadmills use RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A
The CR60 Rehabilitation Treadmills are suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
NOTE Tests verified with shielded input/output cables only.