APPENDIX B
Inspection and Cleaning
If the hospital or institution fails to implement a satisfactory cleaning and inspection schedule for this
equipment, it may result in equipment failure and health hazards.
NOTE:
Only qualified service personnel should repair or replace ELI 250 parts.
Inspecting the ELI 250
Inspect the equipment for the following conditions on a regular basis:
• Examine patient cable, power cable and other accessory cables for obvious damage
(e.g. torn insulation, broken connectors, etc.). Replace cables as necessary.
• No bent prongs or pins on the plugs and connectors. Check the fit of cables with connectors
on the ELI 250 to ensure proper fit.
• All cords and connectors are securely seated in their corresponding connections.
• Examine the equipment for missing screws, cracks or broken areas that might allow unintended
access to internal electronics areas.
Cleaning the ELI 250
Disconnect the power source. Clean the exterior surface of the unit with a damp cloth using a solution of mild
dishwashing detergent diluted in water. After washing, thoroughly dry off the unit with a clean, soft cloth or
paper towel.
Cleaning Print-Head
NOTE:
Do not let soap or water come into contact with the writer, plugs, jacks or vents.
Test Operation
After cleaning and inspecting the ELI 250 proper operation of the unit may be confirmed by using an ECG
simulator to acquire and print a standard 12-lead ECG of known amplitude. Printing should be dark and
even across the page. There should be no evidence of print-head dot failure (e.g. breaks in printing forming
horizontal streaks). Paper motion should be smooth and consistent during printing. Waveforms should appear
normal, with proper amplitude, and without distortion or excessive noise. Paper should stop with perforations
near the tear-bar (indicating proper cue sensor operation).