User Safety Information
(Continued)
• Excessive patient movement could interfere with the operation of the device.
• Proper patient preparation is important to proper application of ECG electrodes and operation of the device.
• There is no known safety hazard if other equipment, such as pacemakers or other stimulators, are used
simultaneously with the ELI 250; however, disturbance to the signal may occur.
• If the ECG amplifier input is out of normal operating range, the display will indicate a lead fail for the
lead(s) where this condition is present and if the signal is being printed, the respective lead(s) will print
out as a square wave.
• As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows:
• Class I equipment or internally powered
• Type CF applied parts
• Ordinary equipment
• Not suitable for use in the presence of flammable anesthetics
• Continuous operation
NOTE:
From a safety perspective, per IEC 60601-1 and derivative standards / norms, this unit
is declared to be “Class I” and uses a three-prong inlet to ensure an earth connection is made
along with mains. The ground terminal on the mains inlet is the only protective earth point in
the unit. Exposed metal accessible during normal operation is double insulated from mains.
Internal connections to earth ground are functional earth.
• The ELI 250 will automatically turn off (blank screen) if the batteries have been severely discharged and
the AC mains is disconnected from the unit.
• After operating the ELI 250 using battery power, always reconnect the power cord. This ensures that the
batteries will be automatically recharged for the next time you use the ELI 250. A light will illuminate,
next to the on/off switch, indicating that the unit is charging. This light will turn off when the battery is
fully charged.
• The ELI 250 is a UL Classified Device:
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE
WITH UL2601-1, IEC60601-1, CAN/CSA CC22.2 No. 601.1,
AND IEC60601-2-25
vi