ELECTROMAGNETIC COMPATIBILITY (EMC)
(Continued)
Table X-1 Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified in the table below. The
customer or the user of the system should assure that it is used in such an environment.
Table X-2 Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the user
of the system should assure that it is used in such an environment.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
x
Emissions Test
Compliance
Compliance Level
Electromagnetic Environment: Guidance
Electrostatic
+/- 6 kV contact
+/- 6 kV contact
Floors should be wood, concrete, or ceramic tile.
Discharge (ESD)
+/- 8 kV air
+/- 8 kV air
If floors are covered with synthetic material, the
IEC 61000-4-2
relative humidity should be at least 30%.
Electrical Fast
+/- 2 kV for
+/- 2 kV for
Mains power quality should be that of a typical
transient/burst
power supply line
power supply line
commercial or hospital environment.
IEC 61000-4-4
+/- 1 kV for
+/- 1 kV for
input/output lines
input/output lines
Surge
+/- 1 kV differential
+/- 1 kV differential
Mains power quality should be that of a typical
IEC 61000-4-5
mode +/- 2 kV
mode +/- 2 kV
commercial or hospital environment.
common mode
common mode
Voltage dips, short
<5% UT
<5% UT
Mains power quality should be that of a typical
interruptions and
(>95% dip in UT)
(>95% dip in UT)
commercial or hospital environment. If the user of
voltage variations
for 0.5 cycle
for 0.5 cycle
the system requires continued operation during
on power supply
40% UT
40% UT
power mains interruptions, it is recommended that
input lines
(60% dip in UT)
(60% dip in UT)
the system be powered from an uninterruptible
IEC 61000-4-11
for 5 cycles
for 5 cycles
power supply or a battery.
Power frequency
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
(50./60 Hz)
characteristic of a typical location in a typical
magnetic field
commercial or hospital environment.
Emissions Test
Compliance
Electromagnetic Environment: Guidance
RF Emissions CISPR 11
Group 1
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to
cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class A
The system is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
Harmonic Emissions
Complies
buildings used for domestic purposes.
IEC 61000-3-2
Voltage fluctuations/flicker
Complies
emissions IEC 61000-3-3