iii)  The test results are displayed at the end of the test.  The results are 

displayed in three windows which may be scrolled with either the 

 or 

 

button.  The first window shows the positive and negative decay times 
(from 1000 volts to 100 volts) for tests number 1 and 2 in the sequence.  
The second window shows decay times for test number 3 and the 
average times for three + and three – decay tests.  The third window 
shows the + and – balance peak voltages and the temperature and 
humidity taken during the test sequence. 

c) 

Press ‘ENTER’ to return to the main menu. 

d) 

Repeat a, b and c to gather a second set of data.  This will be ‘

Test 2

’. 

5.  Run manual decay tests — 

a) 

Press and release ‘+ CHARGE’ and let the first manual decay test run its 
course and display a result.  That manual decay test will be labeled ‘

M1+

’.  To 

return to the main menu, press ‘ENTER’. 

b) 

Press and release ‘+CHARGE’ again and run the test to completion.  This 
second manual decay test will be labeled ‘

M2+

’.  To abort one of these tests at 

any time during the test press the button that got the test started. 

c) 

Return to the main menu.  Press and release ‘

CHARGE’.  This third manual 

decay test will be labeled ‘

M3

’ when completed.  Return to the main menu. 

6.  Run manual balance tests — 

a) 

Press and hold either the 

 or 

 button until Balance test is indicated on the 

display then release.  This will start your fourth manual test.  It will be labeled 

M4B

’ when finished.  Return to the main menu. 

b) 

Repeat for a fifth manual test.  This test will be labeled ‘

M5B

’.  Note that if you 

hold the button down too long to start the test, the test will be nullified.  
Following test ‘

M5B

’, return to the main menu.  The memory usage will be 

displayed as ‘A20%M50%’.  You have conducted 2 out of 10 automatic 
sequences and 5 out of 10 manual tests. Note that if either A or M shows 99%, 
memory is full for that test and another of that test cannot be run until memory 
is cleared. 

7.  At this point, you may elect to shut the equipment down and retire to a different 

location to contemplate the results. 

Test results — 

Press and release the ‘ON-OFF’ button.  You will be prompted for a Y/N? response to the 

ClrData

 (clear data) question.  

NO!!

  Press ‘ENTER’ to accept “N”.  That will take you to the 

main menu while retaining the data. 
 
 
 
 
 
 
 

 

 

Содержание 287B

Страница 1: ...Operator s Manual Model 287B Ionizer Performance Analyzer P N 0340197 287B 1 100 V1 00 08302017 100 Housel Ave PO Box 535 Lyndonville NY 14098 585 765 2254 fax 585 765 9330 monroe electronics com...

Страница 2: ...Check Three steps to get going Page 3 Further familiarization Page 4 Test results Page 5 The Set up Menu Page 6 Maintenance Section 4 Page 7 Precautions Page 7 Cleaning Page 7 Battery Page 7 Calibrat...

Страница 3: ...ions 8 1x4x1 9 inches 206x102x48 mm Weight 1lb 6 oz 0 63 kg Accessories included Carrying case Banana patch cord Alligator clip with boot Battery This manual and a manual supplement What this instrume...

Страница 4: ...e to over 1050 volts During the test the plate will discharge toward zero in the presence of ionization The elapsed time of decay between 1000 volts and 100 volts is displayed USER NOTE For very fast...

Страница 5: ...or message The main menu startup screen will show a blank voltage memory usage status relative humidity present temperature and battery voltage NOTE If the instrument has been used it may display the...

Страница 6: ...with the polarity indicated Thus the 5th test might be labeled M5 3 Shut the instrument off press and release ON OFF You will need at least an ionizer and a ground connection to proceed Further famili...

Страница 7: ...ress the button that got the test started c Return to the main menu Press and release CHARGE This third manual decay test will be labeled M3 when completed Return to the main menu 6 Run manual balance...

Страница 8: ...lect ENTER to execute Unit will return to main display after Enter is executed 3 Auto Seq Auto Sequence Enter presents options or key selects option ENTER accepts choice Default is 4 Tst dly Test dela...

Страница 9: ...e with a clean lint free cloth dampened in the same solution DO NOT use soap or detergent Battery Need for battery replacement is indicated by a warning message whenever the battery voltage persistent...

Страница 10: ...se Min Minutes see AutoShtDwn and The Set up Menu Page 6 Mn Min voltage in balance tests Mx Max voltage in balance tests MnB Balance results for Manual test n where n is test number Mn or Mn Decay tim...

Страница 11: ...Association Inc 7900 Turin Rd Building 3 Suite 2 Rome NY 13440 2069 Phone 315 339 6937 Fax 315 339 6793 eosesd aol com http www eosesd org Other materials listed as Normative References in this docum...

Страница 12: ...epairing or replacing the non conforming goods and or defective work and in accor dance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential R...

Отзывы: