c)  Press and release ‘+ CHARGE’.  This will start a manual test where the test voltage 

is positive. 

(i)  The instrument performs an autozero. 

(ii)  The floating plate is charged to >+1050 volts. 

(iii) The plate begins to discharge.  If allowed to discharge to <+1000 volts the 

timer will start at +1000 volts. 

If discharge fails to reach +1000 volts within 200s, ‘Decay failed’ message 
will be displayed. Same message will appear if the meter fails to decay to 
100 volts within 200 second timer.  Test will not be saved and Automatic 
decay test will be aborted.  

 

(iv) This is a single positive manual decay test.  The test is labeled ‘

M1+

’ 

(

M

anual, Test 

1

+

 [positive]). 

Abort this test by pressing the ‘CANCEL’ button. 

d)  Use the ‘

 CHARGE’ button similarly to perform negative manual decay tests.  Up to 

ten manual decay tests may be run in any polarity order.  The ten tests will be 
labeled 

1

 thru 

0

 with the polarity indicated.  Thus, the 5

th

 test might be labeled ‘

M5

’. 

3)  Shut the instrument off (press and release ‘ON-OFF’).  You will need at least an ionizer 

and a ground connection to proceed. 

Further familiarization — 

In order to become more familiar with the accumulation and presentation of data we will 
perform two automatic test sequences and a few manual tests and then examine the results: 

1.  Set up the ionizer. 

2.  Ground the 287B IPA.  The ground jack is located on the left side of the instrument 

case. 

3.  Access the main menu.  The display should show 0% values for Automatic and 

Manual memory usage. 

If a message prompts you to 

ClrData

 (clear data), 

a)  Scroll to ‘Y’ with either the 

 or 

 button. 

b)  Accept this by pressing ‘ENTER’. 

4.  Run an automatic test sequence — 

a)  Press ‘AUTO’. 

b)  Let the test run its course.  This is ‘

Test 1

’. 

i)  Six decay tests are run.  These are in the order +++



 or +

+

+

 

depending on the current autosequence setup. 

ii) 

A balance test is run.  The duration of the balance test may be any value 
between 10 and 90 seconds or it may be turned off in the SETUP menu.

 

 
 
 
 

 
 

Содержание 287B

Страница 1: ...Operator s Manual Model 287B Ionizer Performance Analyzer P N 0340197 287B 1 100 V1 00 08302017 100 Housel Ave PO Box 535 Lyndonville NY 14098 585 765 2254 fax 585 765 9330 monroe electronics com...

Страница 2: ...Check Three steps to get going Page 3 Further familiarization Page 4 Test results Page 5 The Set up Menu Page 6 Maintenance Section 4 Page 7 Precautions Page 7 Cleaning Page 7 Battery Page 7 Calibrat...

Страница 3: ...ions 8 1x4x1 9 inches 206x102x48 mm Weight 1lb 6 oz 0 63 kg Accessories included Carrying case Banana patch cord Alligator clip with boot Battery This manual and a manual supplement What this instrume...

Страница 4: ...e to over 1050 volts During the test the plate will discharge toward zero in the presence of ionization The elapsed time of decay between 1000 volts and 100 volts is displayed USER NOTE For very fast...

Страница 5: ...or message The main menu startup screen will show a blank voltage memory usage status relative humidity present temperature and battery voltage NOTE If the instrument has been used it may display the...

Страница 6: ...with the polarity indicated Thus the 5th test might be labeled M5 3 Shut the instrument off press and release ON OFF You will need at least an ionizer and a ground connection to proceed Further famili...

Страница 7: ...ress the button that got the test started c Return to the main menu Press and release CHARGE This third manual decay test will be labeled M3 when completed Return to the main menu 6 Run manual balance...

Страница 8: ...lect ENTER to execute Unit will return to main display after Enter is executed 3 Auto Seq Auto Sequence Enter presents options or key selects option ENTER accepts choice Default is 4 Tst dly Test dela...

Страница 9: ...e with a clean lint free cloth dampened in the same solution DO NOT use soap or detergent Battery Need for battery replacement is indicated by a warning message whenever the battery voltage persistent...

Страница 10: ...se Min Minutes see AutoShtDwn and The Set up Menu Page 6 Mn Min voltage in balance tests Mx Max voltage in balance tests MnB Balance results for Manual test n where n is test number Mn or Mn Decay tim...

Страница 11: ...Association Inc 7900 Turin Rd Building 3 Suite 2 Rome NY 13440 2069 Phone 315 339 6937 Fax 315 339 6793 eosesd aol com http www eosesd org Other materials listed as Normative References in this docum...

Страница 12: ...epairing or replacing the non conforming goods and or defective work and in accor dance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential R...

Отзывы: