Section 3 

OPERATION 

Getting Started, a Tutorial

 

 

*Use in conjunction with supplemental guide included on CD 

Confidence Check – Three steps to get going — 

If you received this instrument directly from the factory, a battery was included but not 
installed, memory was cleared and default settings are effective.  The following will 
familiarize you with your new Ionizer Performance Analyzer Model IPA 287B: 

1)  Install battery 

a)  Remove the two crosspoint screws from the battery compartment cover. 

b)  Carefully lift the cover off of the compartment. 

c)  Insert the battery, refer to label for orientation. 

d)  Replace the cover. 

2)  Demonstrate basic functions.  Pressing a single button and releasing within two seconds 

controls basic functions. 

a)  Press and release ‘ON-OFF’.  The display will briefly show information about the 

instrument and then the battery voltage.  A low battery condition will present an error 
message. 

The 

main menu

 startup screen will show a blank voltage, memory usage status, 

relative humidity, present temperature and battery voltage. 

NOTE —

 If the instrument has been used it may display the option of clearing the 

memory.  Do so by scrolling to “Y” with 

 

or

 

 

key then press ‘ENTER’. 

b)  Press and release ‘AUTO’.  This will start the first automatic test sequence: 

i) 

The instrument performs an autozero. 

ii) 

The floating plate is charged to >+1050 volts. 

iii) 

The plate begins to discharge.  If allowed to discharge to <+1000 
volts the timer will start at +1000 volts. 

If discharge fails to reach +1000 volts within 200s, ‘Decay failed’ 
message will be displayed. Same message will appear if the meter 
fails to decay to 100 volts within 200 second timer.  Test will not be 
save and Automatic decay test will be aborted.  

 

iv) 

This test is the first automatic decay test.  The first test in an 
automatic sequence is always positive.  The test is labeled ‘

11+

’ 

(Automatic Test Sequence 

1

, Test Number 

1

 [in a sequence of 6 

tests], and the test voltage is 

+

 [positive]). 

 

Abort this test by pressing the ‘AUTO’ button again (‘CANCEL’).

 

 
 
 
 
 
 

Содержание 287B

Страница 1: ...Operator s Manual Model 287B Ionizer Performance Analyzer P N 0340197 287B 1 100 V1 00 08302017 100 Housel Ave PO Box 535 Lyndonville NY 14098 585 765 2254 fax 585 765 9330 monroe electronics com...

Страница 2: ...Check Three steps to get going Page 3 Further familiarization Page 4 Test results Page 5 The Set up Menu Page 6 Maintenance Section 4 Page 7 Precautions Page 7 Cleaning Page 7 Battery Page 7 Calibrat...

Страница 3: ...ions 8 1x4x1 9 inches 206x102x48 mm Weight 1lb 6 oz 0 63 kg Accessories included Carrying case Banana patch cord Alligator clip with boot Battery This manual and a manual supplement What this instrume...

Страница 4: ...e to over 1050 volts During the test the plate will discharge toward zero in the presence of ionization The elapsed time of decay between 1000 volts and 100 volts is displayed USER NOTE For very fast...

Страница 5: ...or message The main menu startup screen will show a blank voltage memory usage status relative humidity present temperature and battery voltage NOTE If the instrument has been used it may display the...

Страница 6: ...with the polarity indicated Thus the 5th test might be labeled M5 3 Shut the instrument off press and release ON OFF You will need at least an ionizer and a ground connection to proceed Further famili...

Страница 7: ...ress the button that got the test started c Return to the main menu Press and release CHARGE This third manual decay test will be labeled M3 when completed Return to the main menu 6 Run manual balance...

Страница 8: ...lect ENTER to execute Unit will return to main display after Enter is executed 3 Auto Seq Auto Sequence Enter presents options or key selects option ENTER accepts choice Default is 4 Tst dly Test dela...

Страница 9: ...e with a clean lint free cloth dampened in the same solution DO NOT use soap or detergent Battery Need for battery replacement is indicated by a warning message whenever the battery voltage persistent...

Страница 10: ...se Min Minutes see AutoShtDwn and The Set up Menu Page 6 Mn Min voltage in balance tests Mx Max voltage in balance tests MnB Balance results for Manual test n where n is test number Mn or Mn Decay tim...

Страница 11: ...Association Inc 7900 Turin Rd Building 3 Suite 2 Rome NY 13440 2069 Phone 315 339 6937 Fax 315 339 6793 eosesd aol com http www eosesd org Other materials listed as Normative References in this docum...

Страница 12: ...epairing or replacing the non conforming goods and or defective work and in accor dance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential R...

Отзывы: