Section 1 

SPECIFICATIONS: 

 

Charger:

 

1100 volts, selectable polarity 

Fieldmeter: 

Accuracy

 

5% of reading, 

2% typical  

Zero Drift 

<

4V in 90s, 

2V typical

 

Trip points:

 

Fixed 1000V and 100V 

Charge plate:

 

1.7”x4” (

10.6 inches periphery) 

Capacitance  

25pF 

Self-discharge 

<10% of full scale in 200s

 

Display:

 

2x16 LCD, >5 updates per second 

Timer 

200.0s maximum, 0.1s resolution 

Fieldmeter

 

1250V, 1V resolution 

Sensors: 

Humidity 

5% typ. from 10% to 90%RH @25ºC 

Temperature

 

2ºC typical

 

Oper. temp.:   

25ºC, 

10ºC

 

Battery:

 

9 volt NEDA #1604A or equivalent alkaline (>40 hour life or >1300 
charge cycles). Longer life may be achieved by using 9-volt lithium. 

Dimensions:

 

8.1x4x1.9 inches (206x102x48 mm) 

Weight:

 

 

1lb, 6 oz. (0.63 kg) 

 
 

Accessories included: 

Carrying case 

Banana patch cord 

Alligator clip with boot 

Battery 

This manual and a manual supplement 

 
 

 

What this instrument does — 

  Model 287B IPA meets and far exceeds the basic requirements of ESD Association 

(Draft) 

Standard Practice 3.3 for the Protection of Electrostatic Discharge 

Susceptible Items – Periodic Verification of Air Ionizers

  Model 287B IPA performs manual or automatic decay and balance tests for periodic 

verification of ionization equipment and stores the results and averaged decay times for 
up to ten manual tests and up to ten complete automatic test sequences. 

  Provides real-time measurement and display of temperature and relative humidity. 

  Performs self-tests. 

  Battery powered and portable, sets up in seconds. 

 
 

 
 
 
 
 
 

 
 

Содержание 287B

Страница 1: ...Operator s Manual Model 287B Ionizer Performance Analyzer P N 0340197 287B 1 100 V1 00 08302017 100 Housel Ave PO Box 535 Lyndonville NY 14098 585 765 2254 fax 585 765 9330 monroe electronics com...

Страница 2: ...Check Three steps to get going Page 3 Further familiarization Page 4 Test results Page 5 The Set up Menu Page 6 Maintenance Section 4 Page 7 Precautions Page 7 Cleaning Page 7 Battery Page 7 Calibrat...

Страница 3: ...ions 8 1x4x1 9 inches 206x102x48 mm Weight 1lb 6 oz 0 63 kg Accessories included Carrying case Banana patch cord Alligator clip with boot Battery This manual and a manual supplement What this instrume...

Страница 4: ...e to over 1050 volts During the test the plate will discharge toward zero in the presence of ionization The elapsed time of decay between 1000 volts and 100 volts is displayed USER NOTE For very fast...

Страница 5: ...or message The main menu startup screen will show a blank voltage memory usage status relative humidity present temperature and battery voltage NOTE If the instrument has been used it may display the...

Страница 6: ...with the polarity indicated Thus the 5th test might be labeled M5 3 Shut the instrument off press and release ON OFF You will need at least an ionizer and a ground connection to proceed Further famili...

Страница 7: ...ress the button that got the test started c Return to the main menu Press and release CHARGE This third manual decay test will be labeled M3 when completed Return to the main menu 6 Run manual balance...

Страница 8: ...lect ENTER to execute Unit will return to main display after Enter is executed 3 Auto Seq Auto Sequence Enter presents options or key selects option ENTER accepts choice Default is 4 Tst dly Test dela...

Страница 9: ...e with a clean lint free cloth dampened in the same solution DO NOT use soap or detergent Battery Need for battery replacement is indicated by a warning message whenever the battery voltage persistent...

Страница 10: ...se Min Minutes see AutoShtDwn and The Set up Menu Page 6 Mn Min voltage in balance tests Mx Max voltage in balance tests MnB Balance results for Manual test n where n is test number Mn or Mn Decay tim...

Страница 11: ...Association Inc 7900 Turin Rd Building 3 Suite 2 Rome NY 13440 2069 Phone 315 339 6937 Fax 315 339 6793 eosesd aol com http www eosesd org Other materials listed as Normative References in this docum...

Страница 12: ...epairing or replacing the non conforming goods and or defective work and in accor dance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential R...

Отзывы: