Installing a Detection Cartridge
There are three (3) slots available in the detection cartridge drawer. You can install a
detection cartridge into any available slot.
To install a detection cartridge:
1. On the touch screen on the front panel of the instrument, touch the detection
cartridge drawer open/close button.
CAUTION!
To prevent damage to the installed detection cartridges and the
instrument, do not manually slide the detection cartridge drawer in or out
when the instrument is powered on or when one or more detection
cartridges are installed in the drawer.
Figure 1-16: Detection Cartridge Drawer Open
2. Select the slot or slots on the detection cartridge drawer for the detection cartridge.
3. Remove the red cap from the detection cartridge before installation, if supplied
with the detection cartridge. The Transmitted Light (TL) Detection Cartridge does
not require a red cap.
4. Position the two small holes and the connector pins on the detection cartridge over
the holder pins and connector for the detection cartridge slot.
Note:
Some detection cartridges occupy more than one slot.