SpectraMax MiniMax Imaging Cytometer Installation Guide
10
5024111 C
To attach the cytometer to the microplate reader:
1. Place the foam pad from the shipping container on the work bench on one side of
the microplate reader.
CAUTION!
If the SpectraMax i3x Instrument has previously been in use,
remove all microplates and detection cartridges from the instrument and
install the transport locks on the microplate drawer and detection cartridge
drawer. See the user guide for the SpectraMax i3x Instrument.
2. Use two people to carefully place the microplate reader on its side on top of the
foam pad.
WARNING! PINCH HAZARD.
There is limited clearance between the side
of the microplate reader and the work bench.
3. Use the provided 2.5 mm hex key to remove the four screws (G) that secure the
optics-access cover (F) to the bottom of the microplate reader.
Item
Description
F
Optics-access cover
G
Screws (4)
Figure 1-4: Optics Access Cover on the Bottom of the Microplate Reader
4. Remove the optics-access cover (F) from the microplate reader and store the
optics-access cover and screws in the original shipping container for the cytometer.
5. Grasp the foam block in the optics-access opening and then gently pull the foam
block out. Store the foam block in the original shipping container for the cytometer.