SpectraMax MiniMax Imaging Cytometer Installation Guide
22
5024111 C
6. Connect one end of a 2 meter USB cable (S) supplied with the cytometer to a USB
3.0 port in the PCI card on the rear of the computer, and then connect the other end
of the USB cable to the USB port labeled
COMPUTER USB 1
(T) on the rear of the
cytometer.
Note:
You must use the 2 meter USB cables supplied with the cytometer to
connect the instrument to the computer.
7. Connect one end of a 2 meter USB cable (U) supplied with the cytometer to a USB
3.0 port in the PCI card on the rear of the computer, and then connect the other end
of the USB cable to the USB port labeled
COMPUTER USB 2
(V) on the rear of the
cytometer.
Note:
The USB port labeled
COMPUTER USB 2
is the connection for the
camera. It must be connected to a USB 3.0 port in the PCI card on the rear
of the computer.
For SpectraMax
I3 and MiniMax
Figure 1-15: USB 3.0 Ports in the PCI Card
8. If required, connect the power supply to the cytometer port on the rear of the
cytometer labeled
Alternative Power Input
, and then connect the other end of the
power supply to a grounded electrical wall outlet. The alternative power supply is
required only with the SpectraMax i3 Instrument. It is not required for the
SpectraMax i3x Instrument.
9. Set the power switch on the rear of the instrument to the On position and wait for
the instrument to complete its initialization routine.
You might be prompted to install USB drivers, which must be done before you start
the software.
CAUTION!
The instrument can be damaged if the transport locks are not
removed before the instrument is powered on. See
10. Start the SoftMax Pro Software.