SpectraMax MiniMax Imaging Cytometer Installation Guide
20
5024111 C
Item
Description
1
Power switch
2
Off
3
On
Figure 1-13: Power Switch
To connect the cytometer cables:
1. Make sure that the instrument and host computer are placed on a dry, flat work
area with sufficient space for the two devices and the required cables. To ensure
sufficient ventilation and provide access for disconnecting power from the
instrument, maintain a 20 cm to 30 cm (7.9 in. to 11.8 in.) gap between the rear of
the instrument and the wall.
2. Make sure that the power switch on the rear of the microplate reader is in the
Off position.
3. Turn on the power to the host computer.
4. Install the SoftMax Pro Software on the computer. See the SoftMax Pro Software
installation guide or user guide.
The instrument is supported by SoftMax Pro Software version 6.3 or later. The
SoftMax Pro MiniMax Imaging Edition
must be installed to run the
SpectraMax MiniMax Imaging Cytometer.
To prevent data loss, turn off all sleep and hibernation settings for the hard disk,
the CPU, and the USB ports. Also, disable automatic Windows Updates. You can
update Windows manually when the instrument is not being used by the software.
See
Required Computer Settings on page 17
.
Note:
Do not start the software until after the cables have been connected
and power applied to the instrument. It might be necessary to install USB
drivers before starting the software.