5090339 A
Packing the Instrument
The original packaging is designed to protect the instrument during shipment and storage. You
must always pack the instrument before you ship it or store it.
To pack the instrument:
1. Make sure the plate drawer chamber and filter cube chamber are empty.
2. Power off the instrument.
3. Disconnect the power cord and USB cable from the rear of the instrument.
4. Locate the transport lock that you stored in the accessories case.
Item
Description
1
Screw #1 fastens the transport lock to the internal frame of the instrument (install last)
2
Screw #2 fastens the transport lock to the plate drawer
3
Screw #3 fastens the transport lock to the plate drawer
4
Plate drawer
5
Plate chamber door in open position
6
Plate drawer transport lock
5. Manually open the plate chamber door, gently slide the plate drawer out of the instrument,
and place the transport lock on the end of the of the plate drawer.
6. Use the hex key (included in the accessory case) to tighten screws #2 and #3 until the
transport lock is attached to the plate drawer.
7. Gently push the plate drawer into the instrument so that screw #1 aligns with the hole in the
instrument frame. Hold the plate drawer open until you complete the next steps.
8. Tighten screw #1 in the upper-right corner of the transport lock to secure the plate drawer
to the instrument frame.
9. Route the yellow tab connected to the transport lock so that it passes over the top of the
plate chamber door when the door is closed.
10. Gently close the plate chamber door.