Chapter 2: Setting Up the Instrument
5090339 A
19
Installing Filter Cubes
Filter cubes ship with the instrument in a specialized case. Each case can contain up to four
filter cubes. Store filters and filter cubes in a protected, dust-free area, preferably in the supplied
case.
CAUTION! Handle filter cubes by the stem only. Do not touch filter surfaces with fingers.
Fingerprints left on filters negatively affect measurement results.
You insert the filter cube into the filter chamber on the front of the instrument. Each filter cube
has an NFC tag that the instrument reads. The NFC tag contains the information the instrument
uses for reads. Filter information displays in the Instrument Information dialog in the software.
To install filter cubes:
1. Turn on the instrument power switch located on the rear of the instrument. Wait for the
instrument to complete its diagnostic check and the indicator light to turn green.
Note: You can install a filter cube with the instrument powered off, but the instrument
must be powered on and connected to the computer in order for the SoftMax Pro
Software to recognize the filter cube you install.