Maintenance
30
SkanWasher 400 User Guide
Rinse Outlet Channels
If it is determined that the aspirate probes (longer probes) are clogged, follow
this procedure after performing the Clogged Probes procedure to dislodge
particles.
Step 1
Open the outlet channel seal block to get access to the wash head
outlet channel (see Figure 11).
Remove two hex wrench screws using the hex wrenches available
in the tool kit.
Pull out the wash head block to expose the outlet channels.
Step 2
Observe for particles in outlet channel.
Step 3
If particles not removed, and one probe can be identified to be
clogged, replace probe.
Or
If particles not removed, please consult Technical Support to have
outlet channel replaced.
Figure 11: Wash head outlet channel seal block