QPix 450 or QPix 460 Colony Picking System User Guide
68
5031498 G
To adjust the test image:
1. From the Receptacle list, select the number of the receptacle to view and then click the
frame that you want view in the receptacle image below the list. The frame is in red.
2. Under Acquisition, adjust the Exposure and Gain settings.
In the Exposure (ms) field, enter the number of milliseconds (from 10 to 1000) to expose
the sample to light when acquiring the image.
Adjust the Gain to increase the image signal without changing the light exposure.
After making adjustments to the Exposure or Gain settings, click Grab Image to apply the
new settings to the test image.
3. Select the Detection tab.
Содержание QPix 450
Страница 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Страница 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Страница 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Страница 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Страница 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Страница 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Страница 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Страница 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Страница 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Страница 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Страница 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...