FilterMax F3 and FilterMax F5 Multi-Mode Microplate Readers User Guide
42
5016208 D
Physical Specifications
Item
Description
Environment
Indoor use only
Power requirements
100 VAC to 240 VAC ±10%, 3 amps, frequency range: 50/60 Hz
Dimensions
39 cm W x 24 cm H x 58 cm D
(15.4 in. W x 9.5 in. H x 22.9 in. D)
Weight
25 kg (55.1 lbs)
Power disconnect and
ventilation clearance (rear)
20 cm to 30 cm (7.9 in. to 11.8 in.) between the rear of the
instrument and the wall
Ambient operating
temperature
15°C to 40°C (59°F to 104°F)
Ambient storage temperature
-5°C to 40°C (23°F to 104°F) continuous;
-20°C to 50°C (-4°F to 122°F) transient (up to 10 hours)
Humidity restrictions
15% to 75% (non-condensing) at 30°C (86°F)
Altitude restrictions
Up to 2000 m (6562 ft)
Air pressure restrictions
54 kPa to 106 kPa (7.8 PSI to 15.4 PSI)
Sound pressure level
Maximum sound pressure: 75 dBA
Maximum sound pressure at one meter: 68 dBA
Installation category
II
Pollution degree
2
Fuses
250 V, 3 amp, 5 x 20 mm, slow-blow,
UL recognized/CSA/VDE
Data connection
One (1) RS-232, 9-pin serial port
Physical Specifications For The FilterMax F3 and FilterMax F5
Item
Description
Temperature range
3°C (5.4°F) above ambient to 45°C (113°F) for visible range
(>359 nm)
4°C (7.2°F) above ambient to 45°C (113°F) for UV (deuterium
lamp)
Displayed
1°C (1.8°F) increments
Accuracy
±1°C (1.8°F)
Uniformity
<1.5°C (2.7°F)
Temperature Control For The FilterMax F5