Chapter 4: Maintenance and Troubleshooting
5016208 D
29
Item
Description
1
Power Switch
2
Power Port
3
Serial Port
To replace the fuses:
WARNING! HIGH VOLTAGE
Power off the instrument and disconnect the
power cord before you do maintenance procedures that require removal of a
panel or cover or disassembly of an interior instrument component.
1. Power off the instrument.
2. Unplug the power cord from the power port.
3. Use a small flat head screwdriver to gently press on the carrier release tab and then pull
the fuse carrier to remove it from the instrument.
Item
Description
1
Fuse carrier
2
Carrier release tab
Release Fuse Carrier
4. Gently pull the old fuses from the carrier by hand.
5. Place new fuses into the carrier by hand.
6. Press the fuse carrier into the instrument until the carrier snaps into place.
7. Plug the power cord into the power port.
8. Power on the instrument.
Note:
If the instrument still does not power on after you change the fuses, contact
Molecular Devices technical support. See
.