background image

 

 
 

MOLECULAR DEVICES > CLONEPIX 2 

ROBOT MANUAL 

 
 
 
 

 

 

 

 

Molecular Devices > 21 of 39  

 

General Maintenance 

Regular Maintenance 

Daily 

 

Check that the robot interior is clean and free from any obstruction. 

 

Ensure sterile water (feed) bottle is connected to the system fluid supply line. Empty the 
waste bottle every time that a new sterile water bottle is connected. 

 

If the system has been sanitized, purge the system fluid supply line on first use to 
remove air bubbles that will have developed overnight. If there is doubt about sterility, 
run the 'Prepare for Pick Run' routine which includes a purge of the system fluid supply 
line with 70% ethanol or Sterilizing Agent (K8080). Sterilizing Agent (K8080) cannot be 
used in North America. This should be done no more than once a week. 

 

During the purge, check that there are no air bubbles in the supply line coming from the 
sterile water bottle (indicates a fault in the connection), and that there are no leaks 
around the picking head (indicates a blocked pin). If there is a leak, or if a tube detaches 
from the head, replace blocked pin with a new one or follow the pin cleaning procedure 
below. 

 

Fill ethanol wash bath with 70% Ethanol. For systems fitted with a replenishing ethanol 
wash bath, check that there are no kinks in the supply/waste tubing and then run 
through with 70% ethanol. Check that the bath fills with ethanol and that the waste 
ethanol flows freely into the waste bottle. On first use, it may be necessary to run 
through several times. 

 

Weekly 

 

In any week that ClonePix 2 is not used, it is important to purge the system fluid supply 
line with sterile water, and run 70% ethanol through the replenishing ethanol wash bath 
(if fitted). Failure to keep the pumps regularly primed may lead to system malfunction. 

 

Autoclave the picking pins. 

 

Check operation of interlock switch and emergency stop. 

 

Check the compressed air filter for signs of moisture. If necessary, push the drain 
upwards to force moisture out. 
 

Monthly 

 

Sanitize picking head and silicon supply pipes (picking tubing). 

 

Sanitize source plate illumination cover. 

 

Annually 

 

Preventative Maintenance Service – only to be carried out by Molecular Devices approved 
personnel. 

 

As Required 

 

If the robot is used in particularly dusty conditions, the HEPA filter or pre-filter should be 
checked and may need to be replaced more frequently than recommended (by Molecular 
Devices approved personnel).  

Содержание ClonePix 2

Страница 1: ...07MAN1186 B ECO 4100 ClonePixTM 2 ROBOT MANUAL...

Страница 2: ...10 Robot Enclosure 11 Installation 12 Layout of ClonePix 2 12 Source and Destination Plate Stackers 13 Connecting the Wash Bottles 14 Connecting the Ethanol Bottles 14 Camera Alignment 15 ClonePix 2...

Страница 3: ...d Sterile Water Supply 24 Incoming Compressed Air Supply 25 Automated Picking Pin Cleaning 25 Sterility Validation Tests 26 Changing the Picking Head 27 Remove Head 27 Replace Head 27 Changing or Repl...

Страница 4: ...Devices 4 of 39 Table of Figures Figure 1 ClonePix 2 electrical connections 8 Figure 2 ClonePix 2 PC connections 9 Figure 3 ClonePix 2 layout and dimensions 12 Figure 4 Stacker orientation 13 Figure 5...

Страница 5: ...imaging using a choice of five filter sets Colonies selected on the basis of size shape and proximity to neighbors Colonies ranked based on fluorescent intensity of the colony or the secreted product...

Страница 6: ...pth x 1495mm height Weight Robot 350 kg maximum Compressor 60 kg Compressed Air Supply Minimum Pressure 6 bar Minimum volume 80L min Operating Environment Indoor use only Temperature 10 C to 40 C Humi...

Страница 7: ...ater IEC Output PC IEC Output Monitor Fuses Input F3 T10A Robot Halogen Heater Output F2 T5A Computer and Monitor Compressor 230V AC 50Hz 3 4A USA Japan Voltage 100 125V AC 45 66 Hz single phase Power...

Страница 8: ...g instructions 2011 The year of manufacture is given on the serial number label which is located on the right side of the unit European Economic Area EEA This mark on the product indicates compliance...

Страница 9: ...MOLECULAR DEVICES CLONEPIX 2 ROBOT MANUAL Molecular Devices 9 of 39 PC Connections Figure 2 ClonePix 2 PC connections Mains Input Monitor Camera Ethernet Mouse USB Keyboard Printer...

Страница 10: ...yan 4 X4991 standard CloneDetect FITC Fluorescein GFP AF488 470 Blue 3 535 Green 3 X4992 YFP 500 Cyan 2 550 Yellow 2 X4993 CloneDetect 549 R phycoerythrin Rhodamine AF546 530 Green 1 590 Red 1 X4994 C...

Страница 11: ...d exhaust fan This creates a horizontal laminar flow of sterile air across the picking area of the bed of the machine whilst the front door is closed The main features are Class 100 type enclosure Off...

Страница 12: ...Molecular Devices 12 of 39 Installation Note Installation is to be undertaken only by Molecular Devices approved personnel Layout of ClonePix 2 Front view Top view Figure 3 ClonePix 2 layout and dimen...

Страница 13: ...automatically fed onto the bed individually and the lid is removed as the plate is pushed through the lid lift mechanism The lid is replaced as the plate returns through to the cassette Figure 4 Stack...

Страница 14: ...ake sure that they are marked so that the tubes can be re connected correctly The outlets on the machine are in the same order as the bottles on the wash station Figure 5 Wash bottle connections Conne...

Страница 15: ...formed before ClonePix 2 is dispatched and will be checked by a Molecular Devices approved engineer when the robot is installed If the actuator is subject to a knock or a crash picking accuracy may be...

Страница 16: ...ne should be properly balanced on the forks prior to lifting Important note Do not use any part of the exterior bodywork to lift the instrument as this may cause irreparable damage External Covers War...

Страница 17: ...lamp an efficient source of germicidal radiation Emergency STOP Button The location of the Emergency Stop button is shown in Figure 7 Page 15 Striking the Emergency Stop button will immediately stop...

Страница 18: ...ll not exceed 70db measured at a distance of 1 meter Service and Maintenance It is strongly recommend that maintenance be carried out regularly by a Molecular Devices approved service engineer Mainten...

Страница 19: ...n left on the bed of the instrument Check that all motor tracks are free of obstruction Check that there are no obstructions to head movement Check that the main robot axes XY are roughly in the centr...

Страница 20: ...x 2 application by selecting Exit from the File menu on the main setup screen Close down Windows Click the Start menu at the bottom of the screen then click Shut Down Wait for computer to switch off c...

Страница 21: ...pin cleaning procedure below Fill ethanol wash bath with 70 Ethanol For systems fitted with a replenishing ethanol wash bath check that there are no kinks in the supply waste tubing and then run thro...

Страница 22: ...air flow For more thorough cleaning use the following procedure 1 Remove picking head from ClonePix 2 see instructions below and remove pins and picking tubing 2 Submerge head into a beaker of 70 Etha...

Страница 23: ...w 2 Make a 2 solution 50mls of aQu Clean Pin Cleaning Solution K2505 in a small plastic beaker 3 Place all the items in the solution and sonicate for 10 minutes 4 Rinse thoroughly in de ionized water...

Страница 24: ...onditions 121 C at 15 psi 103 kPa for 20 minutes Cleaning of Instrument Interior For efficient decontamination of pathogenic micro organisms all non removable parts within the ClonePix 2 top cover sho...

Страница 25: ...pplied oil free compressed air unit draws air from the local environment and filters it to 0 02 microns using a 3 stage filtration process Automated Picking Pin Cleaning The reusable Picking Pins are...

Страница 26: ...fter this period there was no contamination on any of the agar plates A general purpose medium which supports the growth of fastidious and nonfastidious microorganisms 2 Mammalian cell picking sterili...

Страница 27: ...e right hand side to secure the head Fit the picking tubing to the tube connectors starting at the back and working forward making sure that each tube is fitted to the correct connector Follow the ins...

Страница 28: ...lated with eight X4961 F1 Pins X4942 Additional picking head populated with eight X4962 F2 Pins X4949 Additional picking head without pins X4961 F1 ClonePix picking pin 0 40mm diameter Single pin X496...

Страница 29: ...CIF Complex Initiation Factor 10000U 1ml K8205 CloneDetect human IgG Fc Chain specific FITC label Sterile azide free 10000U 1ml K8206 CloneDetect human IgG Fc Chain specific FITC label Sterile azide...

Страница 30: ...20 CloneMatrix optimized for CHO cells 6 x 40ml 6 x 100ml final media volume CloneMedia XPMedia Cat No Description Pack Size K8610 CloneMedia semi solid media for hybridomas myelomas 1 x 90ml K8600 Cl...

Страница 31: ...W1105 PetriWell 6 Plate Black with clear base Polystyrene Non TC treated 1 40 W1107 PetriWell 6 Plate Black with clear base Polystyrene Collagen coated 1 40 W1109 PetriWell 6 Plate Black with clear ba...

Страница 32: ...in cleaning solution 1L K8080 Sterilizing Agent 1L per Sachet 10 K8085 Sterilizing Agent 1L per Sachet 50 K8150 CaliBeads Fluorescent beads for ClonePix 200 m diameter Pan wavelength 100ml K8010 Adher...

Страница 33: ...ystem Purge the supply line Check Z height of pin Not depositing Pin blocked run Purge option or replace pin Dispense volume too low No liquid flow to Wash Station System fluid supply bottle is empty...

Страница 34: ...ht option normally used for configuring the robot and for visualising colonies during picking using LEDs located under the source plate being imaged EQUIGLASS A high quality polymer used as the base m...

Страница 35: ...ng suspended colonies F2 pins have an internal tip diameter of 700 m and are for picking adherent colonies monolayers PRIMARY PIN The picking pin nearest the front of the robot All other pins are alig...

Страница 36: ...in some older systems WASTE BOTTLE 4 liter bottle used to collect purged fluid 1 liter in some older systems WHITE LIGHT Full spectrum LED illumination of which there are two options Transillumination...

Страница 37: ...trical supply 7 8 9 10 Emergency 15 17 19 G Germicidal radiation 17 H Head 19 I Initial Operation 19 Installation 11 12 16 Isolator switch 20 L Lifting 16 19 Liquid spillage 16 M Maintenance 18 Maximu...

Страница 38: ...MOLECULAR DEVICES CLONEPIX 2 ROBOT MANUAL Molecular Devices 38 of 39 Wash station 14 16 Waste bottles 14...

Страница 39: ...t 2012 by Molecular Devices New Milton Ltd All rights reserved No part of this publication may be reproduced stored in a retrieval system or transmitted in any form by any means electronic mechanical...

Отзывы: