Reliable liquid dispensing
at an affordable price
7
stalling. Replace the tubing if necessary. Frequency of replacement will be dependent
on duty cycle, reagents and maintenance.
RECOMMENDED DMC-B SYRINGE SERVICE PROCEDURE
Reagent to use: Isopropyl alcohol (IPA) 70%
Weekly
Empty the reagent being used. Put inlet Cannula(s) into IPA. Prime for 10 cycles.
Ideally the power to the dispenser can be turned off at the bottom of the fill stroke. This
will give the syringe full contact with the IPA. Leave overnight. Next day, using the
empty command, return the IPA to its container for reuse.
Monthly
After the weekly IPA rinse empty the system. Initiate the empty command again and
turn power off at the bottom of the fill stroke. Undo the syringe plunger thumbscrew and
remove the plunger.
Note: This does not require the glass barrel to be removed.
Dip the white plunger tip into IPA and reinsert the plunger into the barrel. By hand move
the plunger up and down within the bottom 1 inch of the barrel several times. Reinstall
the plunger onto its mounting and replace the retaining thumbscrew. Repeat on all
syringes sets.