GENERAL SYSTEM INFORMATION
1. Power must be off when changing syringes or valves to prevent possible injury in the
advent of accidental triggering.
2. When positioning the inlet and outlet fluid lines avoid any sharp bending that may
cause kinking.
3. Take care when installing the threaded line retainer fittings to the valve ports to
prevent cross threading and possible damage to the valve. Fittings should be firmly
finger tightened only.
4. Do not attempt to internally service the rotary shear valves. Any indication of
disassembly will void the warranty.
5. Handle syringes with care. Do not run syringes dry for more than a couple of cycles.
Doing so can cause accelerated wear of plungers.
6. Use the inlet filter system at all times. Clean at regular intervals.
7. For best results clean and rinse the entire system after each use as described in the
maintenance section.
8. The program memory is nonvolatile and will retain data in power off or power failure
conditions.
DISPENSER SETUP
Refer to figures 1 – 3 for visual assistance with the following instructions.
1. Connect power cord to rear of dispenser
2. Power up dispenser. The
On
switch is located in the rear of the dispenser on the
lower right.
3. Park the dispenser’s drive arm by pressing the [
Mode]
key once to access
dispenser’s Prime function. Then press [
+ / Start]
key. Turn off the power when the
dispenser’s drive arm is at or near the bottom of its stroke.
4. Install the syringe – See steps 4-6 of “Syringe Replacement” on page 8
5. Install inlet manifold – Remove the bottom screws from valves 4 & 6 and replace
with standoffs. Remove the thumbscrews from the ends of the standoffs and mount
the inlet manifold. Replace the thumbscrews – finger tight.