P/N 330654-E 01/2018
COPYRIGHT ©
Dansensor A/S
38
User Guide
EN
Lippke 4000/4500
Pressure drop [%]
(Stop-Threshold)
If there is a problem with false burst detections you can increase the
pressure drop value. Also check that the prefill pressure is high
enough.
Recommended setting: 10%
With oxygen measurement *
Requires a PacCheck 3XX - If checked the O
2
level is measured.
Limit oxygen [%]
(O2 Limit)
Upper limit for acceptable O
2
measurement.
With CO2 measurement *
Requires a PacCheck 3XX - If checked the CO
2
level is measured.
Limit CO2 [%] *
Upper limit for acceptable CO
2
measurement.
Displayed limits in chart:
Pressure, upper limit in chart
[mbar] *
Upper pressure value for the [mbar] axis in the resulting graph chart
Pressure, lower limit in chart
[mbar] *
Lower pressure value for the [mbar] axis in the resulting graph chart
Limits:
Limit + [mbar]
(Upper-Limit)
Upper limit for an ok package. The package should burst below this
limit.
Limit - [mbar]
(Lower-Limit)
Lower limit for an ok package. The package should burst above this
limit.
Содержание Dansensor Lippke 4000
Страница 1: ...P N 330654 E 01 2018 Package Test System Lippke 4000 4500 User Guide EN...
Страница 3: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 2 User Guide EN Lippke 4000 4500...
Страница 7: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 6 User Guide EN Lippke 4000 4500...
Страница 11: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 10 User Guide EN Lippke 4000 4500...
Страница 17: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 16 User Guide EN Lippke 4000 4500...
Страница 51: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 50 User Guide EN Lippke 4000 4500...
Страница 75: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 74 User Guide EN Lippke 4000 4500...
Страница 77: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 76 User Guide EN Lippke 4000 4500...