![Mocon Dansensor Lippke 4000 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/mocon/dansensor-lippke-4000/dansensor-lippke-4000_user-manual_1816853038.webp)
Dansensor A/S
COPYRIGHT ©
P/N 330654-E 01/2018
Lippke 4000/4500
EN
User Guide
37
Measurement accessory kits
The following measurement accessory kits can be used for a
Burst test
:
•
Stand, basic
(with any test head and needle or tube adaptor, with or without filter)
•
Fixture f. ASTM restraining
(with any test head and needle or tube adaptor, with or without filter)
•
PPC 300 II
(with or without restraining plate)
• Tube adapter
• IV-Bag adapter
• Valve Test Unit (VTU)
• Handheld needle, mono
• Handheld, needle, twin
Burst test settings
For a Lippke 4500 the test parameters are set in the PC program and for a Lippke 4000 the parameters
are set on the device display.
* = parameter only available in the Lippke 4500 PC Software
Names in parenthesises are the corresponding Lippke 4000 parameter names
Parameters:
Date last modification *
Filled out automatically
User *
Filled out automatically
Name
(Name)
Test name
Name 1 *
Can be used to give additional information about the test
Name 2 *
Can be used to give additional information about the test
Pressure, rising rate
[mbar/sec]
(Pressure rate)
During the measurement the pressure in the package is increased
linearly at the set rate from the pre-fill pressure until the max.
pressure
Package type
(Package Type)
- Normal
This is the recommended standard setting. Should
be used whenever possible.
- Porous
Use for porous packages like Tyvek.
- Sensitive
Use this setting if the package bursts in the form of
a small hole and the normal setting is having
problems detecting the burst.
The sensitive setting changes the burst limit from a
% of the pressure to a % of the valve setting.
Pre filling pressure [mbar]
(Fill pressure)
If a package for example changes shape when it is filled, then set the
prefill pressure so that this will happen during the prefill time. This
avoids false burst detections.
Maximum pressure [mbar]
(Max Pressure)
The maximum pressure the package is filled to. If the package
doesn't burst then the test stops then the max pressure is reached.
Содержание Dansensor Lippke 4000
Страница 1: ...P N 330654 E 01 2018 Package Test System Lippke 4000 4500 User Guide EN...
Страница 3: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 2 User Guide EN Lippke 4000 4500...
Страница 7: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 6 User Guide EN Lippke 4000 4500...
Страница 11: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 10 User Guide EN Lippke 4000 4500...
Страница 17: ...P N 330654 E 01 2018 COPYRIGHT Dansensor A S 16 User Guide EN Lippke 4000 4500...