Figure 12 shows the program window with the settings for eight modules. The eight channels of

the addressed rack case are listed in rows. In the column 

Input

 you can select the input type. The

types Charge, IEPE and Voltage are available. Under 

Gain

 the amplification is set for each mod-

ule. The gain can be 0.1 (only for charge), 1, 10, 100 and 1000. Under 

Integrator / Highpass

you can select acceleration from 0.1 or 3 Hz, velocity or displacement. Four 

Lowpass

 filters of

0.1, 1, 10 and 50 kHz are selectable. Transducer 

Sensitivity

  is entered as 4 digit number with

decimal point. The decimal point is mandatory. It must not be before the first digit. Valid exam-

ples are “0.123”, “1.234”, “12.34”, “123.4” or “1234.”. 

Module name

 is a 20 character text for

the identification of the channel. Only letters and numbers are allowed. Letters are converted

into upper case. Module names with less than 20 characters are filled with spaces on the left

side.
The settings are transferred to the M72 immediately after entering. Valid entries are confirmed

in the 

Status

 column by 

OK

.  Invalid entries are indicated as 

ERROR

.

Please note that settings transferred via the digital interface are not stored in the non-volatile

memory of the M72 modules. This means that the M72 modules will be reset to the previous

condition which was set via the front panel if the power supply is interrupted. If you want to

keep your settings after switching off the device please click the button “Remanent Memory”.

This will store all settings in the non-volatile memory of the M72 modules.
The program also indicates type and version of the plugged in modules.
Click 

Key Lock

 to disable the front panel keys of all modules. The button becomes yellow when

the keys are locked. The key lock can be released by clicking the button again or by switching

the power supply off and on. 
To save the entered settings to a file click 

Save Settings

. Select a folder and enter a file name.

The program will automatically append the extension “.xml”.
You can reload the saved settings by clicking 

Load Settings

. Select the “.xml” file and click OK.

The loaded settings will be transferred to the M72 modules immediately.
The program is based on LabView 2014. Metra provides the source code on demand.

19

Figure 12: Program window with loaded settings

Содержание M72A1

Страница 1: ...nual Manfred Weber Metra Mess und Frequenztechnik in Radebeul e K Mei ner Str 58 D 01445 Radebeul Phone 49 351 836 2191 Fax 49 351 836 2940 Email Info MMF de Internet www MMF de Charge IEPE Conditione...

Страница 2: ...The latest version of this document can be downloaded from http www mmf de product_literature htm Manfred Weber Metra Mess und Frequenztechnik in Radebeul e K Jan 17...

Страница 3: ...ff the IEPE Power Supply 8 5 Amplifier 8 6 Transducer Sensitivity M72Ax M72S1 9 7 Level Indicators 10 8 Filters and Integrators 10 8 1 Low Pass Filter 10 8 2 High Pass Filter 11 8 3 Integrators 11 9 K...

Страница 4: ...The letter x in this manual stands for the channel number of the versions M72A1 M72B1 M72A3 and M72B3 2 Figure 1 M72A1 front panel Figure 2 M72A1 rear panel...

Страница 5: ...oners provide the following functions Adaptation of the sensor signal and sensor supply Amplification High and low pass filtering for example anti aliasing filter Integration of the sensor signal for...

Страница 6: ...o measure velocity and displacement re spectively A second amplification level follows subsequently Due to the distributed amplification before and after the filter there is an overload reserve which...

Страница 7: ...nel measuring systems In most cases you will find a superimposed 50 60 Hz or 100 120 Hz voltage on the measuring signal One reason for this effect may be that the transducers are ground referenced not...

Страница 8: ...ower supply 4 1 Charge Input Capacitive signal sources usually piezoelectric sensors with charge output are connected to the charge input The input stage is a capacitive feedback amplifier which trans...

Страница 9: ...current must be fed into the measuring line and simultaneously separated from the subsequent amplifier stages The flow of constant current into the sensor it is indicated by the permanently lit white...

Страница 10: ...100 1000 mV pC IEPE mode 1 10 100 1000 times or 0 20 40 and 60 dB The gain can be selected using buttons Gain and Gain After switching the input the lowest amplification is automatically configured Af...

Страница 11: ...he sensitivity is read automatically and appears briefly on the display The most recently entered sensitivity can be viewed at any time by a short press of or It is also saved when the instrument is s...

Страница 12: ...nds For display values of over 80 the gain should be decreased and for under 6 it should be increased to yield an undistorted low noise signal 8 Filters and Integrators 8 1 Low Pass Filter To eliminat...

Страница 13: ...o tating machinery Single integration of vibration acceleration generates velocity whereas double integration generates displacement The HP Integrator button activates the integrators Single integrati...

Страница 14: ...litudes due to their damping behavior As a result of this the modulation declines as does the signal to noise ratio As a guide value single integration should only be applied to frequencies up to 1000...

Страница 15: ...rface see Chap ter 11 10 Function Check and Calibration We recommend calibrating your instrument at two year intervals Inside the housing there are no calibration points All configuration is carried o...

Страница 16: ...l functions of the amplifier modules are controlled via the USB port at the rear of the rack enclosure The power supply socket is also located at the rear side of the rack enclosures It is a circular...

Страница 17: ...s located on the internet site http www mmf de software_download htm m72 Unzip and save both of the enclosed files to a directory on your computer This directory will need to be entered when windows a...

Страница 18: ...meters n CR Send example B1BMEASURING POINT NO 1 RMD set default parameters command acknowledgment a CR RME save all settings to non volatile memory from version 001 005 command acknowledgment a CR RM...

Страница 19: ...t invalid parameters n CR Send example 00S101 2 RMX read settings Answer aaaaa hhh vvv a device type M72R1 or M72S1 nnnnnnnnnnnnnnnnnnnn h hardware version 3 numbers IiGgHhLlSsssss v software version...

Страница 20: ...nder Programs Metra Radebeul on VS1x The pro gram window opens as shown in Figure 11 The rack case M72R8 or M72S8 should now be connected to the PC and the device driver in stalled as described in sec...

Страница 21: ...column by OK Invalid entries are indicated as ERROR Please note that settings transferred via the digital interface are not stored in the non volatile memory of the M72 modules This means that the M72...

Страница 22: ...at modulation 10 of the full scale value at the center of the frequency band Low Pass Filter 3dB 0 1 1 10 50 kHz 4 pole Butterworth damping 75 dB decade High Pass Filter 3 dB 3 Hz 2 pole Butterworth 4...

Страница 23: ...egment LED display Power Supply 7 to 28 VDC 250 60 mA per channel plug connection DIN 45323 Mains Plug Adapter Input range 100 240 VAC 50 60 Hz with 2 pole plug available types EU standard or AU UK US...

Страница 24: ...rranty does not apply to any improper or inadequate maintenance or modification and operation outside the product s specifications Shipment to Metra will be paid by the customer The repaired or replac...

Отзывы: