#RMLl
:
select low pass
l: 0: 0.1 kHz; 1: 1 kHz; 2: 10 kHz; 3: 50 kHz
command acknowledgment:
/a<CR>
error at invalid parameters:
/n<CR>
Send example:
#B5L3
(50 kHz)
#RMO
:
overload condition (from version 001.005)
Answer:
0
no overload since last #O command
/a
1
at least one overload occurred since last #O command
/a
#RMSsssss
:
enter transducer sensitivity
s: sensor sensitivity, 4 digits with the decimal point after 1
st
to 4
th
position
command acknowledgment:
/a<CR>
error at invalid parameters:
/n<CR>
Send example:
#00S101.2
#RMX
:
read settings
Answer:
aaaaa hhh.vvv
a: device type („M72R1“ or M72S1“);
nnnnnnnnnnnnnnnnnnnn
h: hardware version (3 numbers);
IiGgHhLlSsssss
v: software version (3 numbers)
n: device name (20 characters)
i: input (1 number)
g: gain (1 number)
h: high pass / integrator (1 number)
l: low pass (1 number)
s: sensitivity (4 numbers and decimal point)
Answer example:
M72S1 001.002
CHARGE AMPLIFIER
I1G1H1L3S1.000
/a
#RMY
:
Read calibration values
Answer:
ccccc
c: 12 calibration values, 5 digits each
...
(7 x charge; 4 x voltage, 1 x zero corr.)
ccccc
Answer example:
10001
10010
09981
09878
09999
10021
09580
10120
10002
10045
09876
00002
/a
17