MKS Instruments UK Ltd
Microvision2 Hardware Manual – SP101015.102 August 2010
15
4.9 Digital I/O Connector
See Interfacing Notes on Page 16 before making connections to this interface
A standard 25-way D-type socket used for interfacing with external equipment. This may be already in use
depending on the type of RGA system purchased, otherwise it is used to connect accessories such as a
Remote Vacuum Controller.
The Digital I/O connector can also be used to provide alarm output signals and process trip signals.
This interface offers the following functionality:
16 configurable I/O lines
+3.3V power output fused at 120mA, fuses are self resetting
±
15V power output fused at 120mA, fuses are self resetting
+24V power output fused at 120mA, fuses are self resetting
Note: The total power consumption on each rail (+3.3V,
±
15V & +24V) for both the Analog and Digital I/O
ports must not exceed 120mA.
Pin Function
Description
1
Multiplier Trip
Dedicated to Multiplier Trip
2 PA1
3 PA3
4 PA5
5 PA7
6 PB1
7 PB3
8 PB5
9 PB7
10 0V
Digital
11 0V
Analogue
12
-15V
-15V supply output, fused at 120mA
13
+15V
+15V supply output, fused at 120mA
14 PA0
15 PA2
16 PA4
17 PA6
18 PB0
19 PB2
20 PB4
21 PB6
22
+3.3V fused
+3.3V supply output, fused at 120mA
23 Not
Connected
24 GND
25
+24V
+24V supply output, fused at 120mA