Environmental Requirements
Chapter Two: Installation
10
Environmental Requirements
Follow the requirements listed below when installing and using an 1100 Series MFC:
1. The normal operating temperature should remain between 15° and 40° C
(59° and 104° F).
2. The maximum gas inlet pressure is 150 psig.
3. The power input required is ±15 VDC (±5%) @ 300 mA.
A. The maximum voltage/current at startup is ±15 VDC (±5%) @ 300 mA.
B. The typical steady state voltage/current should be ±15 VDC @ 180 mA.
4. The warm-up time is 30 minutes (for warm-up time on 10 sccm units, consult the
factory).
5. For optimal performance, the instrument should remain powered at all times.
For additional MFC requirements refer to Appendix A, Product Specifications, page 39.