How To Zero the Type 127/128 Transducer
Chapter Four: Operation
20
Lowest Suggested Pressure Reading
Full Scale Range
Lowest Pressure Reading
0.1 Torr
5 x 10
-5
Torr
1 Torr
5 x 10
-4
Torr
2 Torr
1 x 10
-3
Torr
10 Torr
5 x 10
-3
Torr
100 Torr
5 x 10
-2
Torr
1000 Torr
5 x 10
-1
Torr
5000 Torr
2.5 Torr
10,000 Torr
5 Torr
15,000 Torr
7.5 Torr
20,000 Torr
10 Torr
25,000 Torr
12.5 Torr
Table 4: Lowest Suggested Pressure Reading
Lowest Suggested Pressure for Control
Full Scale Range
Lowest Pressure Reading
0.1 Torr
5 x 10
-4
Torr
1 Torr
5 x 10
-3
Torr
2 Torr
1 x 10
-2
Torr
10 Torr
5 x 10
-2
Torr
100 Torr
5 x 10
-1
Torr
1000 Torr
5 Torr
5000 Torr
25
Torr
10,000 Torr
50
Torr
15,000 Torr
75 Torr
20,000 Torr
100 Torr
25,000 Torr
125 Torr
Table 5: Lowest Suggested Pressure for Control