Chapter Two: Installation
Electrical Information
15
Interface Connector
Zero Pot
Zero
R
Type 12X
BARATRON
Pressure
Transducer
+15VDCXXXmA
Figure 5: Top View of a Type 127/128 Transducer
Electrical Information
The Type 127 transducer requires an external power source capable of supplying
±
15 VDC,
±
5%,
at 250 mA minimum. The Type 128 transducer requires 500 mA. Noise and ripple should be
less than 20 mV peak-to-peak. Any readout device may be used which has input capabilities of 0
to 10 VDC and impedance greater than 10k ohms.
Note
The ground of any external power supply and readout should be the same
as the sensor ground (chassis ground), to minimize any possible ground
loops which can affect the performance and stability of the system.