Appendix A: Product Specifications
Electrical Specifications
55
Appendix A: Product Specifications
Electrical Specifications
CE Compliance
Electromagnetic Compatibility
1
Machinery Directive
EMC Directive 89/336/EEC
Directive 89/392/EEC
Control Lines
OPEN and CLOSE Valve control lines
(Activated by an active low TTL)
Hi input maximum voltage is 30 Volts and drawing less than
200
µ
A of leakage current
Input Power
Minimum =+14 VDC to m30 VDC @ less than 5
Watts
If a transducer is connected to the 153/1253, its power can be
supplied by the same external supply used by the 153/1253,
or by a separate external supply.
Outputs
Opto-coupled Limit Switch (L.S.)
One for Open, one for Close. 30 V maximum, 10 mA
maximum
Power Loss
Unit saves calibration constants and valve position
Pressure Input Signal
0 to +10 Volts DC (normally)
Can be calibrated to any full scale voltage from 1 to 10 Volts
RS-232
Bi-directional communication
Input - set point and learn commands (in 0.1% increments)
Output - (on request) actual position, pressure (in 0.1%
increments)
“Out of Sync” - Valve Slipped
Set Point Input Signal
0 to +10 Volts DC (normally)
Can be calibrated to any full scale voltage from 1 to 10 Volts
1
An overall metal braided shielded cable, properly grounded at both ends, is required during use.
Содержание 1253D
Страница 3: ...115714 P1 Rev D 3 99 MKS Type 153D and 1253D Throttle Valve Controllers...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 50: ...Valve Slippage Signal Chapter Three Overview 40 This page intentionally left blank...
Страница 64: ...General Information Chapter Five Maintenance 54 This page intentionally left blank...