23
Electromagnetic Flow Meter
FW: 841012/842009
Potential Equalization and Grounding
Type 7100/7200/7600 in Conductive Pipes
Conductive pipes
Equalization connection
Soft gasket
Ground connection.
NB! The flow sensor must be connected to an effective ground
connection, and the wire dimension must be at least 1.5 mm
2
.
Type 7100/7200/7600 in Non-conductive Pipes
Ground connection.
NB! The flow sensor must be connected to an effective ground
connection, and the wire dimension must be at least 1.5 mm
2
.
Non-conductive pipes
Potential equalization connection
Soft gasket
Grounding ring (Min. thickness = 5 mm.)