Operation
4.3 Operating the emergency stop switch
In an emergency, you can stop operating table from moving by pressing the emergency stop switch.
WARNING
The emergency stop switch must be used only in an emergency.
Operating in an emergency
1.
Press the emergency stop switch.
The buzzer sounds and the operating table stops.
Canceling operations
1.
After the operating table stops, turn the emergency stop
switch in the direction of the arrow to cancel the emergency
stop switch.
The buzzer stops sounding.
CAUTION
To reset the operating table to the original
position in an emergency where, for example, an
operator's hand is caught in a gap of the operating
table, press the switch on the control unit to move
the table toward the reverse direction.
NOTE
The emergency stop switch
is located at the left side of
the auxiliary switch.