Page 30
SECTION II OPERATION
WARNING
x
Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes
are set before transferring the patient.
x
Exercise caution with the table top ro-
tated 90° to the base since an improp-
erly distributed patient load may cause
the table to be tipped over.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and car-
diac defibrillator monitors. Improper op-
eration procedures might cause a burn
on an operators or patient, and also
cause the malfunction of the product.
WARNING
x
When a patient is put on a mattress and
his/her position is set up, a patient shall
be set up to more than 1 cm apart from
a side rail so that a patient dose not
touch on side rails (metal).
x
The mattress should be attached in
order not to come off while being in use.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table are
dependent on the use of the original
pad set which was furnished with the
table or an alternate approved re-
placement.
2-13. Positioning
The use of certain optional accessories available
from MIZUHO further extend the positioning capa-
bilities of the 6702 Tables. Refer to the optional ac-
cessories manual or contact your MIZUHO repre-
sentative for further details.
WARNING
Certain accessories may limit weight
capacities. Check with your MIZUHO
representative.
WARNING
MIZUHO Products are guaranteed for
proper performance with the use of
genuine MIZUHO accessories.
Accessories and products not furnished
by MIZUHO have not been tested for
proper performance and safety. Such
applications or use are at the discretion
of the user to ensure patient and staff
safety.
A patient's position setup is performed under the
condition that a mattress is equipped to the upper
surface of table top.
1. Attach a mattress to table board.
A mattress which is made by MIZUHO and is
exclusive use of this table is used. (Refer to
the following Notes)
2. Put a patient on the table.
3. Set up the patient's position in accordance
with the purpose of the operation. (Figure
2-24)