SECTION II OPERATION
2-1. Electrical Power
The 6702 table will operate on either 100 - 240 VAC
or battery power.
WARNING
Prior to operating the table, observe all
table caution labels and review the
SPECIAL USER ATTENTION section in
the front of this manual.
WARNING
Possible explosion hazard exists if table
is used in the presence of
FLAMMABLE ANESTHETICS.
NOTE
An equalization terminal is located un-
der the main power panel. This is pro-
vided as an alternate pathway to re-
duce the risk of static shock hazards.
Always follow recommended grounding
procedures to ensure patient and staff
safety.
2-2. AC 100 - 240 V Operation
Use the following procedures to operate the table on
100 - 240 VAC power.
a.
Make sure the Power cord is securely attached to
the table. To install the power cord, align the cord
connector with the base connector, insert the cord
and twist clockwise to lock the cord into the con-
nector. See figure 2-1. Plug the cord into a properly
grounded, Hospital Grade, 100 - 240 VAC outlet.
Make sure the power cord is routed to the outlet to
prevent it from being in the way of operating per-
sonnel. To remove the power cord, and pull the cord
connector out of the base connector.
WARNING
To avoid the risk of electric shock, this
equipment must only be connected to a
supply main with Protective earth.
Power Cord has to be used the one
which has a logotype of MIZUHO.
b.
Activate the POWER SWITCH located on the
electrical panel. The switch will illuminate.
POWER CORD
BATTERY INDICATOR
POWER SWITCH
FOOT SWITCH
E
F
MAIN POWER
SWITCH
BATTERY
INDICATOR
EQUALIZATION
GROUNDING TERMINAL
FOOT CONTROL
CONNECTOR
POWER
CORD
Figure 2-1. Electrical Panel
The pendant control buttons and the green AC 100 -
240 V, POWER indicator light located in the upper
right corner of the pendant control will illuminate.
See figure 2-2.
AC100 -240 V POWER
ON INDICATOR LIGHT
(GREEN)
Figure 2-2. Pendant Control
c.
The table is now ready for 100 - 240 VAC opera-
tion.
d.
When operating table is disconnected from
commercial power supply, a power supply cable has
to be pulled out from AC outlet.
WARNING
Please don't position the surgical table
in such way that the operator has diffi-
culty to disconnect the Equipment by
plug or appliance outlet.