Vp+ & Vbw Adjustments... PWB-SCAN
Note:
Measure voltage between the TP and closest Chassis Ground.
1.
Vp+ - Adjust VR1 so the voltage at TP33 is set to the level specified for Vp+ on the Voltage Label +/- 0.5V.
2.
Vbw - Adjust VR2 so the voltage at TP34 is set to the level specified for Vbw on the Voltage Label +/- 0.5V.
Vp+
VR1 TP33
VR2 TP34
Vbw
PWB-SCAN