t
eCHNICal
S
PeCIfICatIoNS
A-1
Appendix A
t
echnIcAl
s
pecIFIcAtIons
In this section, “typical” refers to a specification that is within ±20% of a stated value at room
temperature (25°C/77°F).
VAPR ELECTRODE SPECIFICATIONS
Overall Length:
5.7 – 8.9 in (14.5 – 22.5 cm)
Working Length:
3.7 – 6.5 in (9.5 – 16.5 cm)
Shaft Diameter:
0.06 – 0.18 in (1.5 – 4.5 mm)
Shaft Bend Angle:
0 – 30°
Active Tip Orientation:
0 – 90°
Supplied Sterile and For Single-Use Only
Sterilization Method:
Irradiation
VAPR HANDPIECE AND CABLE SPECIFICATIONS
Overall Length:
10.5 ft (3.2 m)
Sterilization Method:
Steam (134 – 136°C)
Supplied Non-Sterile
20 x Reusable
VAPR GENERATOR SPECIFICATIONS
Dimensions (H x W x D):
3.5 in x 16 in x 14.5 in (9 cm x 41 cm x 36.8 cm)
Weight:
12 lb. (5.6kg) approximately
Transport and Storage Conditions
Ambient Temperature Range:
32 to 122°F (0 to 50°C)
Relative Humidity:
10% to 90%, non condensing
Atmospheric Pressure:
500 to 1060 millibars
Operating Conditions
Ambient Temperature Range:
60 to 104°F (10 to 40°C)
Relative Humidity:
10% to 90%, non condensing
Atmospheric Pressure:
500 to 1060 millibars
Power Supply
Regulation Voltage:
90 – 132V RMS, 198 – 264V RMS
Operating Range:
nominal 100 – 120/220 – 240V RMS 50/60Hz
Inlet Fuses:
Time lag 5A (T5AH 250V)
(VAPR II ONLY)
Inlet Fuses:
Time lag 6.3A (T6A3H 250V)
(VAPR 3 ONLY)
Leakage Currents:
Within limits of Class BF equipment as per IEC 60601-1
Alarm Volume:
Adjustable between 40 dB (minimum) and 65dB
(maximum) at 1m. This is an activation signal only.
Classification:
Electrical: Class 1 ordinary equipment as per IEC60601-1.
EMC: Group 1 Class A as per IEC60601-1-2.