EMC PRECAUTIONS
Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information in the
accompanying documents.
WARNING
Portable and mobile RF communications equipment can affect medical electrical equipment.
WARNING
The use of accessories and cables other than those for which the system was designed can
significantly degrade emissions and immunity performance.
WARNING
Keep the accessory cables away from cables from other electrical equipment. Electrical currents
may be induced in the other equipment causing unintended effects.
WARNING
Do not use a monopolar generator/accessories simultaneously with the VAPR 3 generator.
Activation of a monopolar generator/accessories may cause interference with the VAPR 3
generator resulting in user message changes on the display. Before proceeding with surgery,
confirm proper power settings are displayed on the generator. Ensure the appropriate output
setting is enabled for the desired surgical outcome.
ELECTROSURGICAL SMOKE CAUTION
• Studies have shown that smoke generated during electrosurgical procedures can be
potentially harmful to surgical personnel. Use appropriate surgical masks or other
means of protection.
• It is recommended that a suitable extraction method is employed.
PRIOR TO SURGERY
Operator Safety Warnings
• Electric Shock Hazard: Do not connect wet accessories to the handpiece or generator.
Ensure that all accessories are securely and properly connected.
• Electric Shock Hazard: Do not remove or tamper with the Generator housing. Contact
DePuy Mitek technical service for assistance.
• The power cord must meet all requirements for safe grounding. Do not use extension
cords, multiple point plugs or 2 to 3 pronged adapters.
• Do not reuse or resterilize accessories labeled “SINGLE USE,” as malfunction, injury or
cross-infection may result.
• Non-function of the Workstation may cause interruption of surgery. Ensure that all
installation procedures are followed and all connectors are correctly inserted before
use. A backup system / method should be made available for use.
Operator Safety Cautions
• Inspect the insulation of all cords for cracks, nicks and breaks. Inspect all connectors
for damaged or missing parts.
• Use default power levels to test Electrode performance. Confirm proper default power
settings with package insert information before proceeding with surgery.
• Accessories labeled “REUSABLE” must only be processed according to the
recommended procedures provided in this manual.
• Provide as much distance as possible between the electrosurgical generator and other
electronic equipment (such as monitors) because an activated electrosurgical generator
may cause interference with them.