![Misonix neXus Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/misonix/nexus/nexus_instructions-for-use-manual_1806345053.webp)
Instructions For Use – neXus Ultrasonic Surgical Aspirator
Doc #: 100-10-1000 rev A
p 52
Fuse Replacement (The fuse holder is located on the console rear)
Disconnect the power
cord from the rear of the
console
Remove Fuse Holder
Pinch the tab on
each side of the
fuse holder.
Pull fuse holder out.
Replace Fuses and
reinsert holder into
socket
Replace both fuses as specified above and insert holder
into
the socket. A click shall be heard.
Connect power cord to
the rear of the console
neXus Splash screen shall appear followed by the setup screen
Table 10.2 Console Fuse Replacement