Instructions For Use – neXus Ultrasonic Surgical Aspirator
Doc #: 100-10-1000 rev A
p 25
6.2.
Initial Setup
6.2.1 Connect the Power Cord. The rear of the console features receptacles for the power cord. Insert the power cord into
the receptacle so that it is firmly seated.
6.2.2 Install the IV Pole into the IV Pole receptacle. The IV Pole is keyed so that it can be installed in only one direction,
with the irrigation bag hanging away from the top of the console.
6.2.3 Install the Suction Canister Holder. Ensure the Suction Canister Holder is firmly seated so that it can adequately support.
and secure the Suction Canister.
6.2.4 Install the supplied aspiration tubing from the Suction Canister to the Vacuum Filter.
Section 6.3 sets up the irrigation and aspiration lines to the Canister and the IV bag.
NOTE: IV bag should be 1000ml or less
Figure 6.2.1: Console rear
CAUTION
When using optional cart accessory, be sure to align the rubber feet on the bottom of the console with
the indents in the cart base.
CAUTION
Do not transport the neXus console with the IF bag attached to the IV pole.
6.2.5 Obtain the wireless footswitch and place it in the vicinity of the console.
NOTE: The wireless footswitch is factory paired to the console. Confirmation that the footswitch is communicating to the
console can be found in section 6.4.6.
Suction Canister
Holder
Suction Canister
Saline Bag
IV Pole/Holder
Vacuum Filter
Power Cord Inlet and equipotentiality terminal.
Potential equalization connector makes the connected equipment potential equal (ref IEC 60601-1)
Optional Cart
Accessory
(part number 100-80-0000)