Instructions For Use – neXus Ultrasonic Surgical Aspirator
Doc #: 100-10-1000 rev A
p 26
6.3
Power Up and Setup
The front of the console features a receptacle for the handpiece cable. A large color LCD touchscreen provides the user
interface to adjust the main functions of the console. The side of the console features magnetic receptacles where
disposable tubing pucks are seated to orient the tubing correctly to the irrigation pump and aspiration valve.
WARNING
Improper connection of the handpiece cable may present a shock hazard. Confirm that
handpiece connector is dry prior to plugging it in.
CAUTION
The system check should always be done in advance of preparing patient for surgery to minimize
risk to patient in case of system malfunction.
CAUTION
It is strongly advised that a sterile backup handpiece be readily available in the operating room as
insurance any contamination or malfunction of the handpiece used during surgery.
6.3.1 Startup Screen
Plug the power cord into a hospital grade outlet. Plug the other end of the power cord into the rear of the console. Be sure
that during use there is sufficient room necessary to disconnect the cord from the console for mains disconnection.
WARNING
Proper system grounding can only be ensured when an approved, hospital-grade receptacle and matching power
cord are used. To avoid the risk of electric shock, this equipment must only be connected to a supply with protective
earth. Install plug and receptacles as per local regulations before operating the unit. Power cord, plug and receptacle
should be examined to verify that they are in good working condition before connecting the console. Never pull on
the power cord to remove it from the receptacle.
WARNING
Connecting the console to a power outlet with inadequate voltage or frequency may cause the unit to malfunction
or to create a shock or fire hazard. Refer to the back label on the system console for line voltages and frequencies.
The Startup Screen is the first thing you see upon power up. This screen will progress to the Setup Screen within 5
seconds (see section 6.12 for powering down).
Figure 6.3.1 Startup Screen