LB0261 Rev A 01/2018
© Miramar Labs, Inc. 2018, Confidential Information of Miramar Labs, Inc.
Page 32
•
Terminate Treatment Session
You may want to end a treatment session without completing any remaining placements
for this patient. Referring to the instructions in Section 9.4, you should skip the remaining
placements on both axilla (if operating in Guided Mode) and follow the Touchscreen GUI
to complete the cleaning and wrap-up procedures. Alternatively, you can touch the
Gear
Menu Icon
at the top left corner of the screen to display the Gear Menu Drop Down.
Then select the
Stop Session Icon
shown in Section 9.2. This will terminate the
treatment session and return you to the “Welcome” Screen.
•
Emergency stop
To instantly shut-off of the system at any time, press the red Emergency Stop Button on
the Console as shown in Figure 26 and as described in Section 3.1.
Figure 26: Red Emergency Stop Button
•
Total System shutdown
See Section 9.4. After completing a treatment session, turn the Console OFF by pressing
the green Standby button.
9.6.
OPERATION – POST TREATMENT
The miraDry bioTip must only be used for a single treatment session for one patient. This will
avoid biocontamination.
Step 1 Remove bioTip
Remove the bioTip from the miraDry Handpiece by pressing the button on the Handpiece as
shown in Figure 27.
Figure 27: “Remove bioTip” Screen
Dispose of bioTip according to local regulations.