Microwave Energy Level Settings
There are five pre-set microwave energy levels. Levels determine how long (in seconds) the
microwave energy is applied to a particular target site. The System default microwave energy
level is Setting 1 (2.40 sec).
Microwave Energy
Level #
Microwave Energy Delivery
Time (sec)
1
2.40
2
2.55
3
2.70
4
2.85
5
3.00
The energy level is fixed to level 1 for the first few rows of each guided treatment template,
regardless of the selected energy level.
Total # of rows for
template
Treat these rows at
energy level 1
11
4
13
4
15
5
17
6
Touching the button under the lock symbol on the microwave energy level screen will leave the
fixed energy level 1 in the first few rows. You may override the fixed energy level for a single
treatment (patient) by touching the button under the unlock symbol on the microwave energy
screen.
Alternatively, select the gear icon at the top left corner of the screen. The dropdown menu will
appear which includes a lock symbol. Touching the lock symbol will display a popup screen. By
touching the check mark on the popup screen you can unlock the fixed energy levels. Touching
the X button (
) will leave the default fixed energy level 1 in the first few rows.
On the top left corner of the treatment screen, the green energy level icon will be displayed
below the gear icon. The selected energy level will be shown on the left, above the radio button
icon, and the current energy level will be shown to the right (
).
9.3.
OPERATION – PRE-TREATMENT SETUP
You must complete these steps prior to initiating treatment.
Step 1 Attach the Handpiece to the Console
If the miraDry Handpiece has not already been attached, you will be prompted to do so at this
time as shown in Figure 9. Refer to Section 9.1 for attachment instructions. Turn the System
OFF before connecting the Handpiece.
Figure 9: “Attach Handpiece” Screen