Optosense LLC
USER MANUAL
Design department
SMALL-SIZE EXPLOSIVE GAS MEASURING
SENSOR MIPEX-02-X-X-X.1 X (RX)
ESAT.413347.005 UM
Revision 3.0
January 19th, 2017
Page 28 of 46
THE SOLE PROPERTY OF OPTOSENSE LLC. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
APPENDIX B. CONNECTION DIAGRAM
MIPEX-02
Combined input
parameters:
IECEx/ATEX:
U
i
= +5.0 VDC, I
i
= 450 mA,
P
i
= 0.25 W, C
i
= 38.8
μF,
L
i
= 0 mH.
CAN/CSA:
V
max
= +5.0 VDC, I
max
= 450 mA,
P
max
= 0.25 W, C
i
= 38.8
μF,
L
i
= 0 mH.
NET
+
Tx
Rx
–
Intrinsically safe
power supply and
monitoring circuit
EXPLOSION HAZARDOUS AREA
NET
V
dd
TxD output
RxD input
GND
Fig. 13. Connection diagram for MIPEX-02
USB
Pin
1
2
3
4
NET
+5V
D+
GND
D-
Pin
1
2
NET
SPC
GND
V
mes
NET
RxD input
TxD output
R
MIPEX-02
NET
+
–
Tx
Rx
V
dd
MIPEX-02/03
interface board
GND
Fig. 14. Connection diagram for MIPEX-02 via MIPEX-02/03 interface board (ESAT.200200.00) during
span calibration (SPC
– pins for measuring sensor power consumption; V