5.
Press
Off
(Yes) button enter into sub-menu item.
[0 kg] is displayed. [0] is blinking.
6. Empty the weighing pan.
7.
Press
Off
(Yes) button to start zero point calibration.
The display shows the calibration weight value with all digit blinking [001500.0]
(in this example, 150 kg)
[--C--] is displayed while zero reading is stored.
8. For changing the calibration weight value, press the
Unit
(No) button to edit the
value.
The first digit is blinking [_015.000 kg].
9. Press
Off
(Yes) button to accept and highlight the next digit.
The 2nd digit is blinking [0_15.000 kg].
7 Calibration and Adjustment