In the event of an error during the calibration process, [CALE] (calibration error) is
displayed and the process is canceled.
5.4.2
[LIN] linearization
10. Press the
Unit
key to increase the value, or the
Mode
F
key to decrease the value.
11. Repeat until all digits are correct.
The value of the calibration weight is displayed using flashing digits in the
display: [00100.0] (in this example 100.0 kg)
12. If the value of the calibration weight is correct, place the specified weight on the load
plate.
13. Press the
Off
(Yes) key to adopt the second calibration point.
[--C--] is displayed while the calibration weight is being saved.
14. Remove the weight.
The scale is ready for operation.