4. Press the desired reference quantity by briefly pressing the
Mode
F
(decrease in
increments) key or the
Unit
(No) (increase in increments) key in order to switch
between the selection options [10, 20, 50, 100, 200].
5.
Press the
Off
(Yes) key to confirm the reference quantity.
If the load plate is empty, [PUT.PW] (apply reference weight) is displayed in the
display.
6. Place the desired reference quantity on the load plate or in the container and press
the
Off
(Yes) key in order to determine and save the new reference sample weight.
If there is no reference sample weight on the load plate or the weight is less than
2 d, [REF.ERR] (reference error) is briefly displayed in the display. Apply higher
sample weights.
If there are reference sample weights on the load plate, [PW OK] (adopt sample
weight) is displayed in the display for 2 seconds, then [CLR.LIM] (delete limit
values).
1)
7.
In order to specify new limit values, press the
Off
(Yes) key. Or: in order to use the
saved limit values, press the
Unit
(No) key and proceed with step 16.
[SET.LOW] (specify lower limit) is displayed on the scale.
5 Operating
Bench scale Puro®
Minebea Intec
EN-25