5.2.3.2
Check Counting application
Note:
The application must be previously activated in the menu. See Chapter
The Check Counting application can be used to determine whether the quantity of the
sample is within specified tolerance limits.
Example: The limit values for the tolerance limits are specified as 500 items (lower limit)
and 510 items (upper limit).
Release the key. The application is activated.
If a reference sample weight is already saved, [CLR.PW] (delete reference sample
weight) will be displayed.
or
Note:
1)
If [LIM.ERR] (limit value error) is briefly displayed followed by [CLR.LIM]
(delete limit value), then invalid limit values have been defined. Repeat the
setup process.
1.
In order to select the Check Counting application or to restart with new limit values,
press and hold the
Mode
F
key until [CHECK] (checking) is displayed at the lower edge
of the screen with the application icons
and
.
2.
Press the
Unit
(No) key in order to use the saved reference sample weight (PW), and
proceed with step 11.
3.
Press the
Off
(Yes) key in order to delete the saved reference sample weight (PW),
and specify a new sample weight.
[PUT.20] (for example) is displayed on the display.