The following links between the terminal contacts are provided:
①
from + Supply (+V) to + Sense (+S)
②
from - Supply (-V) to - Sense (-S)
+ Meas.
– Meas.
+ Supply
+ Sense
– Sense
– Supply
Ground
analog
bu
rd
gy
gn
GNDA
PR 5900
PR 5900
2
1
3
4
6
5
③
Screen
④
Screen clamping rail or cable gland
⑤
Equipotential bonding conductor
⑥
or
4.6.4.3
Connecting a load cell with a 6-wire cable
Note:
The colors listed here apply for the Minebea Intec load cell and connection cables of
type "PR …"
Color code
bk
=
Black
bu
=
Blue
gn
=
Green
gy
=
Gray
rd
=
Red
wh
=
White
For additional information on the connection of load cells and cable junction boxes,
refer to the corresponding installation manuals.